EasyRapid – Fixed prosthetic solution

All the abutments, analogues and transfers are made of Grade 5 titanium (Ti6Al4V). The fastening screw ELPSC1 is included in the packaging of the abutments. Tighten to 30 N * cm using the torque ratchet wrench ELPSK1 or ELPSK2 with the torque ratchet ELKWD.

Screw-retained abutment for cemented prosthesis

Moncone-avvitabile-per-protesi-cementata

Code Description Colour
ELMAC1 Short Gold
ELMAC2 Medium Blue
ELMAC3 Long Green

Tighten to 30 N*cm using the wrench ELMAK
with the torque ratchet ELKWD.

 

Castable for screw-retained abutment

Calcinabile-per-moncone-avvitabile

Code Description
EDCAM1 Short
EDCAM2 Medium
EDCAM3 Long

 

Analogue for the screwed abutment

Analogo-del-moncone-avvitabile

Code Description Colour
EDMAC1A Short Gold
EDMAC2A Medium Blue
EDMAC3A Long Green

 

Titanium abutment for cemented prosthesis

Moncone-in-titanio-per-protesi-cementata

Code
ELDTA

 

Milling titanium abutment

Moncone-in-titanio-fresabile

Code
EDMFD


 

Angled titanium abutments

Moncone-in-titanio-preangolato

Code Inclination Type Colour
ELA15A 15° A Green
ELA15B 15° B Purple
ELA25A 25° A Blue
ELA25B 25° B Gold


angolati
Note: The difference between abutments type ‘A’ and type ‘B’, consists in a rotation of 22.5 ° of ‘octagon with respect to its axis of reference and that of inclination. This allows you to double your chances of placement of such abutments.
 

Platform switching abutments

Monconi-Platform-Switching

Code Ø Max (mm) Inclination
ELSP45 4.5 -
ELSP4515 4.5 15°
ELSP55 5.5 -
ELSP5515 5.5 15°

 

Shouldered abutments

Monconi-a-spalla

Code Ø Max (mm) Shoulder (mm)
EDS5010 5.0 1.0
EDS5015 5.0 1.5
EDS5020 5.0 2.0
EDS5510 5.5 1.0
EDS5515 5.5 1.5
EDS5520 5.5 2.0

 

Castable abutment

ELCPXI

Code
ELCPXI

 

Implant analogue

Analogo-impianto
Code Colour
EDAN Gold
 

Implant transfer

Transfer-per-impianto
Code Colour
EDTRA Gold
 

Impression transfer technique for Easy Rapid® implant

To transfer the location of the fixture onto a casting, proceed by inserting the implant transfer EDTRA in the implant and secure it with the long fastening screw ELPSC2 included in the packaging of the transfer. Tighten using the screwdrivers ELCSK1 or ELCSK2 or the contra-angle wrench ELPSCAK. Then proceed with the impression using a slotted spoon. Before removing the impression, loosen the fastening screw releasing the transfer. Insert the implant analogue EDAN, assembling it to its transfer EDTRA using the fastening screw previously loosened. Now it’s possible to send the impression to the laboratory.