EasyNeck – Immediate loading solution

All the abutments, analogues and transfers are made of Grade 5 titanium (Ti6Al4V). The fastening screw ELPSC1 is included in the packaging of the abutments. Tighten to 30 N * cm using the torque ratchet wrench ELPSK1 or ELPSK2 with the torque ratchet ELKWD.

Octagonal abutment

Code
ELMEO

Tighten to 30 N*cm using the torque ratchet wrench ELSAPK
with the torque ratchet ELKWD.

 

Castable for octagonal abutment

Code Indication For implants
ENCEOC For circulars Ø 4.75 & Ø 5.25 mm
ENCEOE For mono implants Ø 4.75 & Ø 5.25 mm

 

Esthetical octagonal abutment

Code Height (mm)
ELMEO15 1,5
ELMEO3 3.0
ELMEO4 4.0

 

Titanium cap for esthetical octagonal abutment

Code
ELMEOC

 

Castable cap for esthetical octagonal abutment

Code
ELMEOCC

 

Protection for octagonal abutment

Code For implants
ENMEOH Ø 4.75 & Ø 5.25 mm

 

Analogue for octagonal abutment

Code Colour For implants
ENAMEO Blue Ø 4.75 & Ø 5.25 mm

 

Transfer for octagonal abutment

Code Colour For implants
ENICE Blue Ø 4.75 & Ø 5.25 mm

 

Implant analogue

Codice Colour For implants
ENAN Green Ø 4.75 & Ø 5.25 mm

 

Implant transfer

Code Colour For implants
ENTRA Green Ø 4.75 & Ø 5.25 mm

 

Impression transfer technique for the octagonal abutment

To transfer the location of the octagonal abutment onto a casting, proceed by inserting the transfer for octagonal abutment ENICE and secure it with the long fastening screw ELPSC2 included in the packaging of the transfer. Tighten using the manual screwdrivers ELCSK1 or ELCSK2 or the contra-angle wrench ELPSCAK. Then proceed with the impression using a slotted spoon. Before removing the impression, loosen the fastening screw releasing the transfer. Insert the implant analogue for octagonal abutments ENAMEO, assembling it to its transfer ENICE using the fastening screw which has been previously loosened. Now it is possible to send the impression to the laboratory.

Impression transfer technique for the EasyDip® implant

To transfer the location of the fixture onto a casting, proceed by inserting the implant transfer ENTRA in the implant and secure it with the long fastening screw ELPSC2 included in the packaging of the transfer. Tighten using the screwdrivers ELCSK1 or ELCSK2 or the contra-angle wrench ELPSCAK. Then proceed with the impression using a slotted spoon. Before removing the impression, loosen the fastening screw releasing the transfer. Insert the implant analogue ENAN, assembling it to its transfer ENTRA using the fastening screw previously loosened. Now it’s possible to send the impression to the laboratory.